SITRANS RD200
3
Technical specifications
SITRANS RD200
Mode of operation
Measuring principle Analog to digital conversion
Measuring points
•
1instrument
•
Remote monitoring of 100instruments
with PC and RDsoftware
Input
Measuring range
•
Current
•
4…20mA, 0…20mA
•
Voltage
•
0VDC…10VDC, 1…5V, 0…5V
•
Thermocouple temperature
•
TypeJ: ‑50…+750°C (‑58…+1382°F)
•
TypeK: ‑50…+1260°C (‑58…+2300°F)
•
TypeE: ‑50…+870°C (‑58…+1578°F)
•
TypeT: ‑180…+371°C (‑292…+700°F)
•
TypeT, 0.1°resolution: ‑180.0…+371°C
(‑199.9…+700°F)
•
RTD temperature
•
100Ω RTD: ‑200…+750ºC
(‑328…+1382ºF)
Output signal
Output
•
4…20mA (optional)
•
ModbusRTU
Relays 2SPDTFormC relays, rated 3Aat 30VDC
or 3Aat 250VAC, non-inductive, auto-
initializing (optional)
Communications
•
RS232 with PDC or ModbusRTU
•
RS422/485 with PDC or ModbusRTU
Accuracy
4…20mA optional output ±0.1%FS ±0.004mA
Process input ±0.05% of span ±1count, square root:
10…100%FS
Thermocouple temperature input
•
TypeJ: ±1°C (±2°F)
•
TypeK: ±1°C (±2°F)
•
TypeE: ±1°C (±2°F)
•
TypeT: ±1°C (±2°F)
•
TypeT, 0.1°resolution: ±1°C (±1.8°F)
RTD temperature input
•
100Ω RTD: ±1°C (±1°F)
Rated operating conditions
Ambient conditions
•
Storage temperature range ‑40…+85°C (‑40…+185°F)
•
Operating temperature range ‑40…+65°C (‑40…+149°F)
Design
Weight 269g (9.5oz) (including options)
Material (enclosure)
•
1/8DIN, high impact plastic, UL94V-0, col
or: gray
•
Optional plastic, steel and stainlesssteel
(Type304, EN1.4301) NEMA4 enclosures
Degree of protection Type4X, NEMA4X, IP65 (frontcover); panel
gasket provided
Electrical connection
mA output signal 2-core copper conductor, twisted, shielded,
0.82…3.30mm² (18…12AWG), Belden
8760 or equivalent is acceptable
Electrical connection and relay connection Copper conductor according to local
requirements, rated 3Aat 250VAC
Power supply
Input voltage option1 85…265VAC, 50/60Hz; 90…265VDC,
20Wmax.
Input voltage option2 12…36VDC; 12…24VAC, 6Wmax.
Transmitter power supply One or two isolated transmitter power
supplies (optional)
•
Single power supply One 24VDC ±10% at 200mAmax.
•
Dual power supplies Two 24VDC ±10% at 200mA and
40mAmax.
Technical specifications (continued)
SITRANS RD200
External loop power supply 35VDCmax.
Output loop resistance
•
24VDC, 10…700Ωmax.
•
35VDC (external), 100…1200Ωmax.
Displays and controls
Display
•
14mm (0.56inch) high LED
•
2Xoption for 30.5mm (1.2inch) high,
redLED
•
Numeric range from ‑1999…+9999
•
Four digits, automatic lead zero blanking
•
Eight intensity levels
Memory
•
Non-volatile
•
Stores settings for minimum of 10years if
power is lost
Programming
•
Primary: front panel
•
Secondary: meter copy or PC with
SITRANSRD software
Certificates and approvals CE, UKCA, UL,
C
UL
Options
Enclosures Plastic, steel, and stainlesssteel (Type304,
EN1.4301) NEMA4 and 4X enclosures
Mounting
•
2inch (5.08cm) pipemounting kit
(zincplatedseal)
•
2inch (5.08cm) pipe mounting kit (stain
lesssteel, Type304, EN1.4301)
SITRANS RD200 · 4/2024
© Siemens AG 2024